Respiratory swab on the isothermal test card, step by step

How a clinic runs a canine or feline respiratory panel on the isothermal molecular line, from an eye, nose or throat swab to a printed result on the palm-sized MiniDock. The same handling applies to every card in the line that takes a swab.

Respiratory swab on the isothermal test card, step by step · 3:26 · filmed in the lab

Step by step

The procedure as the video shows it. Tap a time to play from that step.

  1. Lay out the kit next to the MiniDock: reaction card, Releasing Agent with cap, swab and waste bag.
  2. Put the sampled swab into the Releasing Agent tube.
  3. Break the swab at its lowest point and leave the tip in the tube.
  4. Close the tube and shake the Releasing Agent hard for 15 seconds.
  5. Take the reaction card out of its foil pouch.
  6. Add the Releasing Agent to the card slowly, running it down the side of the inlet, until the liquid level is between the two lines.
  7. Close the cap tightly and wait 15 seconds.
  8. Shake the card vertically and quickly 10 times, then check it; a card that fails the check is discarded and a new one prepared.
  9. Open the lid when the MiniDock is ready, insert the card and close the lid to start the test.
  10. In the software, double-click the test in the information list, enter the sample details and click Modify to save them.
  11. Wait for the result, check it and print the report.
  12. Take the card out to return the MiniDock to ready status, put it in the waste bag, seal the bag and discard it.