Catalogue · Isothermal amplification · Operation videos
Feline ascitic fluid for coronavirus, step by step
When a cat has abdominal effusion, the fluid itself can be run on the feline coronavirus card. The video prepares the ascites by centrifugation and shows two ways to take the supernatant into the Releasing Agent, by pipette or by dipping the swab, before the standard card procedure.
Step by step
The procedure as the video shows it. Tap a time to play from that step.
- Add 1.5 to 2.0 mL of ascites to a tube and centrifuge at 3,000 to 5,000 rpm for 5 minutes.
- Turn on the MiniDock and let it preheat.
- Take out a Releasing Agent tube.
- Transfer 50 µL of the supernatant into the Releasing Agent tube, by pipette or by immersing the swab in the supernatant.
- Close the cap and mix well.
- Open a reaction card and add the Releasing Agent slowly, running it down the side of the inlet, until the liquid level is between the two lines.
- Tighten the cap and wait 10 seconds.
- Press down the air bag to expel bubbles from the chambers, then shake the card up and down for 10 seconds.
- Check that all the reagent has dissolved and that bubbles fill no more than a third of any chamber.
- Insert the card and close the lid immediately; the test starts automatically.
- Double-click the test in the list and complete the test information.
- After the test, click Print Report.
- Take out the card, seal it in the waste bag and discard it. A long press on the button switches the MiniDock off.
Cards this procedure applies to
More operation videos
Respiratory swab on the isothermal test card, step by step3:26 · Isothermal amplification
Rectal swab for the gastrointestinal test cards, step by step3:39 · Isothermal amplification
Whole blood on the isothermal test card, step by step3:22 · Isothermal amplification
African swine fever virus on the isothermal test card, step by step2:58 · Isothermal amplification
The immunofluorescence analyser, from set-up to result1:02 · ImmunofluorescenceGet the written procedure with your quotation.
Send the cards or tests you are evaluating. The quotation comes with the package inserts, specifications and certificates, usually within two working days.